product description
Technical Specifications
PARAMETER |
SPECIFICATION |
Product name |
Ethylene Oxide Sterilizer |
Product model |
VEO-H40 |
Product form |
Desktop |
Heating method |
Electric heating |
Power |
1.3KVA |
Power |
AC220V, 50Hz |
Dimensions |
(Length × Width × Height): 670 × 450× 400mm |
Liner size |
(Length × Width × Height): 240 x 470mm |
Effective Volume |
23L |
Weight |
56kg |
Temperature display |
3.5 inch LCD display |
Material of Sterilization Room |
304 anti-corrosion stainless steel |
Outer box material |
ABS plastic |
Temperature during sterilization |
50℃±3℃ |
Working pressure |
-60kpa |
Highest anti-stress value |
-140kpa |
Humidity during sterilization |
45%-80% |
Selected gas |
Pure ethylene oxide gas |
Sterilization time |
0-99 hours adjustable |
Air Replacement |
Automatic ventilation, 5 cycles of ventilation, every 15 minutes |
Full-automatic control system |
One-click start to finish, the whole process is completed automatically |
Sterilization record
|
The system can store 500 pots of sterilization records |
Display status
|
Display operating status, temperature, humidity, pressure, sterilization time, residual gas treatment time, times, steam generator temperature. |
Humidity Sensor |
The built-in humidity sensor automatically detects the humidity in the cavity. |
Pressure Sensor |
The built-in pressure sensor automatically detects the pressure in the cavity |
Temperature Sensor |
The built-in temperature sensor automatically detects the temperature in the cavity, the accurate range is 50℃±3℃. |
Vacuum system |
Use oil-free vacuum pump. |
Seal door |
304 stainless steel aluminum material, thickness ≥ 2mm, pressure resistance 200Kpa. |
Humidification System |
Using pulse steam humidification |
Residual gas treatment device |
The filtration rate of the filter to the particle diameter ≥0.3μm is not less than 99.5% |
Vacuum rate |
Under no-load conditions, the time for the sterilizer to be vacuumed from normal pressure to -60 kPa ≤ 30 min |
Automatic dosing |
Puncture technology built into the sterilization chamber. |
Heating system |
The explosion-proof electric heating device is used in the sterilization chamber |
Multiple protection functions |
With over-voltage and over-temperature protection functions, it is more safe and reliable |
Alarm system |
Have high temperature alarm system, ultra low pressure alarm system |