Features
- 8-inch color TFT, clear monitoring of EtCO2/CO2, respiration rate, SpO2 and Pulse (Optional)
- ARI’s EtCO2 technology of infrared ray gas analysis.
- Temperature, Pressure compensation system and Zero readjustment system, all of them insure the accuracy and stability.
- Balance gas compensation makes minimum influence against other gases.
- KPA,mmHg or % are selectable
- Audio-optical alarm for exceeding limits and Apnea.
- 24 hours trend diagram keeps still available after power-off.
- Large capacity lithium battery for 10 hours UPS
Technical specifications
PARAMETER |
SPECIFICATION |
Measurement |
CO2 Unit: mmHg, kPa, % (selectable) |
EtCO2 Update: Every Breath |
|
Trend Length: 24hours |
|
CO2
|
Measurement Range:0~150mmHg |
Accuracy: EtCO2 range 0-40mmHg±2mmHg EtCO2 range 41-70mmHg ±5% EtCO2 range 71- 100mmHg±8% EtCO2 range 100-150mmHg±10% |
|
PR |
RR Measurement Range: 3~150t/min |
Accuracy:The bigger value of ± 1% or±1t/min 0~150mmHg |
|
Trend length: 24 hours |
|
Alarm |
Alarm Type: Visual and Audible Alarms |
Alarm situations: Check adapter; No Adaptor; Apnea; Low battery; The value of ETCO2/RR exceeds the upper and lower alarm limits. |
|
Apnea:Low battery;the value of ETCO2/PR exceeds the upper and lower alarm limits. |
|
Oximeter Data (Optional) |
Oximetry Saturation Range:0-100% |
Accuracy ±3% when 50-69% ±2% when 70-100% |
|
Trend length: 24 hours |
|
Pulse Rage |
Range:30-250bpm |
Accuracy:The bigger value of ±2bpm or±2%(reading) when 30-250bpm |
|
Trend Length:24 hours |
|
Power |
Voltage: AC100V-250V,50Hz/60Hz |
Battery: DC3.6V/3800mAh |
|
AC Power Consumption:≤5VA |
|
Battery working time: 4 hours after fully charged |
|
Degree of Protection |
IPX32 |
Operating environment |
Working temperature: 5~40°C |
Relative humidity: 30-75% (non-condensing) |
|
Atmospheric pressure:55~115kpa |
|
Dimension |
Battery: Lithium battery, rechargeable. |
Size:70cm(W)*160cm(H)*44cm(D) |
|
Net Weight: 400g (Lithium battery and adult adapter included) |
|