product description
Feature
- 1-inch Six Standard parameters: ECG, RESP, NIBP, SPO2, TEMP, PR/HR
- 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
- Efficient resistance to interference of defibrillator and electrosurgical cautery;
- SPO2 can testing for 0.1% Weak;
- RA-LL impedance Respiration;
- Trend Coexist Display;
- OxyCRG DyNamic View Display;
- Bed to Bed view Display;
- Networking capacity and nurse calling system;
- Options of Internal Recorder, IBP, EtCO2, Central Monitoring System, Touch Screen, VGA, Neonate SPO2
- Built-in rechargeable battery 4400mAh;
- 1″ high resolution color TFT LCD display;
- Large volume of tabular and graphic trends information storage and easy to recall;
- Anti-ESU, anti- defibrillator;
- Capture dynamic waveforms.
Technical Specification
ECG
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection: I, II, III, avR, avL, avF, V,
Waveform: 2 ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL)
Lead selection: I, II, III,
Waveform: 1 ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
Technical Specification
ECG
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection: I, II, III, avR, avL, avF, V,
Waveform: 2 ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL)
Lead selection: I, II, III,
Waveform: 1 ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
Heart Rate
Range: 15 ~ 300 bpm
Accuracy: ± 1% or ± 1bpm, which great
Resolution: 1 bpm
Sensitivity: >200 (uV P-P)
Differential Input Impedance: > 5 MΩ
CMRR
-Monitor: > 105 dB
-Operation: > 105 dB
-Diagnosis: > 85 dB
Electrode offset potential: ±300mV
Leakage Current: <10uA
Baseline Recovery: < 3S after Defi.
ECG Signal Range: ±8 m V (Vp-p)
Bandwidth
-Surgery: 1 ~ 15 Hz
-Monitor: 0.5 ~ 35 Hz
-Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy: ±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available
Review: Available
Respiration
Method: Impedance between R-F (RA-LL)
Differential Input Impedance: >2.5 MΩ
Measuring Impedance Range: 0.3 ~ 5.0Ω
Base line Impedance Range: 0 ~ 2.5 KΩ
Bandwidth: 0.3 ~ 2.5 Hz
Resp. Measuring Range:
-Adult: 0 ~ 120 rpm
-Neo/Ped: 0 ~ 150 rpm
Resp. Alarm Range:
-High alarm range limit: 8~120 rpm
-Low alarm range limit: 6~120 rpm
Resolution: 1 rpm
Accuracy: ±2 rpm
Apnea Alarm: 10 ~ 40 S
NIBP
Method: Oscillometric
Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
120, 180, 240,480 (Min)
Measuring Period in STAT Mode: 5 Min
Pulse Rate Range: 40 ~ 240 bpm
Alarm Type: SYS, DIA, MEAN
Measuring and alarm range
-SYS: 40 ~ 280 mmHg
-DIA: 10 ~ 225 mmHg
-MEAN: 20 ~ 240 mmHg
Resolution Pressure: 1mmHg
Accuracy Pressure Maximum Mean error: ±5mmHg
Maximum Standard deviation: ±8mmHg
SpO2
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified
Actualization interval: about 1 Sec.
Alarm Delay: 10 Sec.
Pulse Rate
Measuring and Alarm Range: 20~250bpm
Resolution: 1bpm
Accuracy: ±3bpm
Temperature
Measuring and Alarm Range: 0 ~ 50°C
Resolution: 0.1°C
Accuracy: ±0.1°C
Actualization interval: about 1 Sec.
Average Time Constant:<10 Sec.
IBP (optional)
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
-ART: 0 ~ 300 mmHg
-PA: -6 ~ 120 mmHg
-CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
-P1/P2: -10 ~ 300 mmHg
Press Sensor
Sensitivity: 5uV/V/mmHg
Impedance: 300-3000Ω
Resolution: 1 mmHg
Accuracy: ±2% or ±1mmHg, which great
Actualization interval: about 1 Sec
Standard Packing List
5-lead ECG cable (including RESP)
SpO2 sensor
NIBP cuff
NIBP extension tube
TEMP sensor
AC power adaptor
Fuse
Operation instruction
Optional Function
Internal Recorder, IBP, EtCO2, Central Monitoring System, VGA