product description
Characteristics
- The steam can’t go through to chamber when the door hasn’t been closed, and the pilot off inform You can process dry operate.
- When the pressure inner over 0.027MPa, the door be locked and can’t be open.
- If you not choose the sterile time and start the machine, buzzer sound.
- If you mistake open the door in the course of start, the jacket value auto off and open the chamber value.
- Lack water the pilot on, heating and sterile pilot off, the buzzer sound and auto release steam.
- Overload and leakage protection, when leakage or over current occur, it auto off the power.
- Temperature controlled not work and the inner pressure over 0.165MPa, the safety value auto start opens.
- The surface of the sterilizer is steel coated, the chamber of the sterilizer is made of stainless steel.
Technical Data
Model technical data |
V-WS-280YDA |
V-WS-200YDA |
V-WS-150YDA |
Sterilizing chamber volume |
280L(φ600×1000mm) |
200L(φ515×1000mm) |
150L(φ440×1000mm) |
working pressure |
0.22 MPa |
||
working temperature |
134℃ |
||
Adjustment of temperature |
40℃ – 134℃ |
||
Time of sterilization |
0-60 min |
||
Time of drying |
0-60 min |
||
Heat average |
≤± 2℃ |
||
Power |
12KW/380V 50Hz |
9KW/380V 50Hz |
9KW/380V 50Hz |
Dimension |
1400×770×1780 |
1400×670×1650 |
1400×600×1300 |
Package |
1680×920×2100 |
1560×820×1850 |
1550×750×1850 |
GW/NW |
465/365 KG |
350/260 KG |
320/240 KG |